Trystar SureImage Power Distribution Unit

Trystar SureImage Power Distribution UnitThe Trystar SureImage Power Distribution Unit, Model 600/M Medical-Grade, three phase, provides power conditioning, isolation, and power distribution for imaging and treatment modalities.

Designed specifically for “Medical OEM” equipment, the SureImage Model 600/M PDU allows OEMs to overcome the challenges of providing power quality to all system components, regardless of voltage requirements and power demand. The 600/M is a single box solution providing voltage conversion, power conditioning, and OEM-specified output distribution – all in a space-saving footprint.

The Model 600/M, PDU includes a single main input circuit breaker, low impedance double-shielded isolation transformer, and surge protection – together with customized controls and distribution options that allow OEM engineers to “build the perfect PDU” for each and every modality.

Trystar SureImage Power Distribution Unit Features

  • Intermittent K(i) to continuous KVA ratio of 3:1 or lower
  • Conservative transformer design approach permits longer operation at intermittent K(i) ratings to meet the momentary power demands of medical equipment
  • Multiple input voltage taps are easily accessed and selected to accommodate any site’s power source
  • Multiple output voltages available at OEM-specified KVA/K(i) levels
  • Custom controls available to facilitate local and remote on/off features, and emergency power (EPO/REPO) shutdown
  • OEM-specified output power distribution
  • Input surge protection
  • Output TVSS and high frequency noise filters with fuse failure indicators
  • Heavy-duty casters for convenient placement and leveling feet and seismic brackets for secure installation

Trystar SureImage Power Distribution Unit Specifications

  • The following example specification is for our 55 KVA/130K(i) rated model. This specification will vary based on OEM-specific requirements.
  • Input Voltage: 200/208/240/380/415/440/460/480/500 VAC, 3 Phase, Delta
  • Input Frequency: 50/60 Hz, ± 3 Hz
  • Input Circuit Breaker: Adjustable electronic circuit breaker, 250 amp frame
  • Controls: 24 VAC shunt trip control of input CB for emergency power off (local and remote) and over temperature shutdown
  • Output Voltages: 480/277 and 400/230 VAC, 3 Phase, Wye
  • Output Power Rating: 55 KVA/130K(i) total
  • Output Impedance: 2% typical
  • Load Regulation: ≤ 2% at each nominal output voltage, no load to full load, continuous KVA rating
  • Electrostatic Shielding: Two (2) full length copper foil shields
  • Common Mode Noise Attenuation: 126 dB
  • Transverse Mode Noise Attenuation: 3 dB down at 1 KHZ, decaying 40 dB per decade
  • Operating Temperature: 0 degrees to 40 degrees C
  • Efficiency: 97 – 98% typical
  • Audible Noise: ≤ 50 dBA at 1 meter
  • Weight: 988 lbs
  • Dimensions: 22″ W x 26″ D x 49.5″ H

For more in-depth specifications, see the data sheet.

For more information about Trystar Power Products,