The Trystar SureImage Power Processor Model 700F/M (60K(i) – 260K(i)), Medical-Grade, is a power conditioning voltage regulator which provides stable and clean power to CT, MRI, PET, Linear Accelerator, X-ray, and other imaging/treatment equipment, ensuring accurate and clear image scans. The Model 700F/M uses a proprietary power sampling technique to ensure that the proper voltage value is always present at the load and that there are no damaging effects of over or undershoot of voltage. The Model 700F/M is provided in a front access only enclosure to minimize footprint, while increasing serviceability and ease of access.
The Model 700F/M has been shake-table tested and each K(i) size is offered with an optional seismic rating. Seismic-rated Model 700F/M units are available with an input nominal voltage of 208VAC, 240VAC, or 480VAC and are provided on floor-mounting channels. These units meet seismic standards which satisfy the California Office of Statewide Health Planning (OSHPD) requirements, as well as CBC 2013, and IBC 2015 standards.
A K(i) rating refers to the intermittent kVA or momentary power demand rating.
Trystar SureImage Power Processor Features
- Highest Level of Image Quality and Diagnostic Reliability
- Optimum Workflow and Patient Satisfaction
- Removal of Voltage Sags and Surges Associated with System Restarts
- Compensation for Generator Power Voltage Drops During Scans
- Ultimate Component and System Reliability
- Permanent Location or Portable (Optional Casters)
- Small Footprint
- High Efficiency
- Medical Rated and Designed
Trystar SureImage Power Processor Specifications
- Front Access
- Output Line Voltage Regulation: ± 2%
- Output Load Regulation: < 2% from typical steady state load to intermittent power demand
- K-Factor Rating: K13
- Output Impedance: 2% typical at rated load
- Common Mode Noise Attenuation: 146 dB
- Transverse Mode Noise Attenuation: 3 dB down at 1KHz, 40 dB down per decade
- Efficiency: 97% typical at full load, continuous rating
- Internal Regulator Bypass
- Standard Surge Protection Device (SPD): Secondary-connected, 40kA per mode; with high frequency filter, status indicator
- Meets ANSI/IEEE C62.41 category B-3 surge capability
- Optional SPD, 100kA per phase, UL 1449 Listed; with LED status indicators
- Optional Input and Output Digital Metering
- Optional Output Circuit Breakers (Up to 2)
For more in-depth specifications, see the data sheet.